Research of CFD Simulative Optimization on Impeller Structure of Axial Blood Pump

YUN Zhong,TAN Jian-ping,GONG Zhong-liang
DOI: https://doi.org/10.3969/j.issn.1001-2354.2006.10.003
2006-01-01
Abstract:High shearing force produced by the shape of interior flow field of axial blood pump is the important factor leading to blood hemolysis.Hence,the fluid kinetic analysis of blood inside the axial blood pump is helpful to optimize the function of blood pump and to reduce the hurt of blood cells.Utilizing the computer this paper solved the RNG k-eequation and carried out simulation analysis on distribution status of flow-line,speed,pressure and stress etc.of blood in the being designed impeller of micro-axial blood pump.The result of study shows that the liquid inside the blood pump has very complicated flowing conditions.For the sake of avoiding the status of flow separation and too big variation of pressure,it is necessary trying to find a way to lower the shearing stress of fluid.For this reason,a corresponding opinion of improvement was put forward on the impeller structure of blood pump and the impeller structure after optimization was presented.
What problem does this paper attempt to address?