Micro-axial Flow Blood Pump′s Blade Flow Field Calculation Using CFD

Chen Xin,Tan Jianping,Yun Zhong
DOI: https://doi.org/10.3321/j.issn:1003-8728.2007.03.008
2007-01-01
Abstract:The fluid dynamic analysis of blood in an artificial blood pump and the calculation of the structure and flow performance of the blade flow field help gain a better understanding of the blood flow of the blood pump to be designed and of the damage of flow field on blood.Using CFD,the paper calculated the blade flow field of a micro-axial flow blood pump,the rotational speed and driving power of the blade when it meets the needs of supplying blood for a normal human body,thus providing theoretical data for the design of magnetic coupled driving.When blood flows through rotating blades,its speed distribution gradients become smaller,allowing it to flow evenly and reducing its damage,indicating that the structure of the blade flow field is basically reasonable.However,there is a high local shear stress area at the blade edge clearance,and because of the difference in local blood pressure,local circumfluence occurs at the outlet of the blood pump′s flow field.
What problem does this paper attempt to address?