The method improvement of determination of Tetracycline Tablets by HPLC

Ping HUANG,Lin GAO,Lei SIMA,Yakun RAO,Xia LIU,Yan ZHENG
DOI: https://doi.org/10.3969/j.issn.1004-2407.2014.03.014
2014-01-01
Abstract:Objective To establish an improved HPLC method for the determination of Tetracycline Tablets .Methods The separa-tion was performed on a Kromasil C18 (250 mm × 4 .6 mm ,5 μm) column ,with the mobile phase of 0 .1 mol · L -1 diammonium hydrogen phosphate-acetonitriles(84∶16) at a flow rate of 1 .0 mL · min-1 .The column temperature was set at 35 ℃ ,and the detection wavelength was 280 nm .Results Tetracycline presented a good linear correlation in the range of 0 .024 0-0 .769 7 mg · mL -1 (r=1 .000) ,the limit of detection was 21 ng · mL -1 ,and the average recovery was 100 .4% ,with RSD of 0 .8% (n=9) .Conclusion This method was simple ,accurate ,reproducible ,and suitable for the quality control of Tetracycline Tablets .
What problem does this paper attempt to address?