Determination of Tetracyclines in Pharmaceutical Wastewater by Solid Phase Extraction-Ultra Fast Liquid Chromatography-Tandem Mass Spectrometry

LI Zai-xing,ZHANG Fei-fei,ZUO Jian-e,ZHANG Jing-xuan,LI Hui,YU Xin,JU Pan-pan
DOI: https://doi.org/10.3969/j.issn.1003-6504.2013.07.028
IF: 11.4
2013-01-01
Environmental Science & Technology
Abstract:A method based on solid phase extraction-ultra fast liquid chromatography-tandem mass spectrometry(SPEUFLC-MS/MS)was developed to determine tetracyclines such as xoytetracycline,tetracycline,chlortetracycline in pharmaceutical wastewater.Water samples were firstly purified through protein precipitation process with 30% zinc sulfate and 20% potassium ferrocyanide,and the tetracyclines was enriched and purified with Oasis HLB SPE,separated through Agilent ZORBAX SB C18 UFLC using 0.1% formic acid solution and acetonitrile as mobile phase,and detected by mass spectrometry in positive ion mode with multiple reaction monitoring using an electrospray ionization interface.The detection limit was 0.02 μg/L.The linear range was 0.010 to 1.0 mg/L,with a correlation coefficient of 0.998 2~0.999 8.The recoveries at the spiked concentration of 0.20,0.40 and 0.60 μg/L ranged from 80.8% to 97.2%,with a relative standard deviation from 5.31% to 9.92%.The procedure would provide a method for determination of tetracyclines in pharmaceutical wastewater.