Radiation dose optimization in temporal bone CT with iterative reconstruction technique

郑慧,李玉华,李惠民,虞肞崴,汪登斌,王利军
DOI: https://doi.org/10.13609/j.cnki.1000-0313.2014.06.016
2014-01-01
Abstract:Objective:To evaluate the feasibility and the optimal parameters for imaging of the temporal bone by using low-dose-256-slice CT with iterative reconstruction technique(IRT)(iDose 3).Methods:One exsomatized cadaveric head was scanned with spiral scanning mode on 256-slice CT scanner from Philips.Three levels of tube tensions from 120kV to 80kV were chosen for the temporal bone scan.At every tube tension,multiple current-time products were used from 250mAs to 70mAs in 30mAs successive interval decrease.Reconstructions were employed by using IRT (iDose Level 3)and edge-enhangcing reconstruction kenel Y-sharp(C).The visibility of 1 6 anatomical landmarks was scored using a five-point-quality rating.The image quality and radiation dose of every level were assessed.Results:The protocols of"100kV+70mAs+iDose level 3+Y-sharp (C)"can clearly depict the normal anatomy,and the radiation dose is lower than the other else, with effective dose of about 0.15mSv.Conclusion:Low-dose-256-slice CT with IRT is feasible for temporal bone scanning with sufficient anatomical detail.The protocols of"100 kV+70 mAs+iDose level 3+Y-sharp(C)"are the optimized ima-ging protocols.