A microfluidic chip for absorbance measurements with long optical path-length based on a liquid-core waveguide technique

WB Du,F Qun,ZL Fang
DOI: https://doi.org/10.3321/j.issn:0251-0790.2004.04.004
2004-01-01
Abstract:In this work, a system for high sensitivity absorbance measurements on microfluidic chips was developed with a detection volume of 240 nL. Long optical path-length (1.7 cm effective path length) was achieved on the chip by using an integrated liquid core waveguide(LCW) silica capillary coated with Teflon AF. The performance of the system was tested by using o-phenanthroline-Fe(II) complex as a model sample with spectrophotometric detection, achieving a detection limit of 8 nmol/L for the complex. The absorbance of samples ranging from 0.03-50 mumol/L is linear(R-2 = 0.9994) and showed good repeatability with a RSD of 0.8% (n = 5). The sensitivity of the chip-LCW system was a factor of 1.7 higher than conventional spectrophotometry using a 1 cm cell. The present system can be applied in continous flow or flow injection analysis systems.