Whole column fluorescence imaging on a microchip by using a programmed organic light emitting diode array as a spatial-scanning light source and a single photomultiplier tube as detector.

Kang-Ning Ren,Qiong-Lin Liang,Bo Yao,Guo-An Luo,Li-Duo Wang,Yu-Di Gao,Yi-Ming Wang,Yong Qiu
DOI: https://doi.org/10.1039/b707118c
IF: 7.517
2007-01-01
Lab on a Chip
Abstract:A novel miniaturized, integrated whole- column imaging detection (WCID) system on a microchip is presented. In this system, a program controlled organic light emitting diode (OLED) array was used as a spatial-scanning light source, to achieve imaging by the time sequence of the excited fluorescence. By this mechanism, a photomultiplier tube (PMT) instead of a charge coupled detector (CCD) can be applied to the imaging. Unlike conventional systems, no lenses, fibers or any mechanical components are required either. The novel flat light source provides uniform excitation light without size limitations and outputs a stronger power by pulse driving. The scanning mode greatly reduced the power consumption of the light source, which is valuable for a portable system. Meanwhile, this novel simplified system has a broader linear range, higher sensitivity and higher efficiency in data collection. Isoelectric focusing of R- phycoerythrin (PE) and monitoring of the overall process with WCID were performed on this system. The limit of detection (LOD) was 38 ng mL(-1) or 3.2 pg at 85 nL per column injection of PE. The system provides a technique for WCID capillary isoelectric focusing (cIEF) on chip and can be used for throughput analysis.