An Integrated Enzyme-Linked Immunosorbent Assay System with an Organic Light-Emitting Diode and a Charge-Coupled Device for Fluorescence Detection.
Hizuru Nakajima,Yukiko Okuma,Kazuhiro Morioka,Mayo Miyake,Akihide Hemmi,Tatsuya Tobita,Masayuki Yahiro,Daisuke Yokoyama,Chihaya Adachi,Nobuaki Soh,Koji Nakano,Shuhua Xue,Hulie Zeng,Katsumi Uchiyama,Toshihiko Imato
DOI: https://doi.org/10.1002/jssc.201100429
IF: 3.614
2011-01-01
Journal of Separation Science
Abstract:A fluorescence detection system for a microfluidic device using an organic light-emitting diode (OLED) as the excitation light source and a charge-coupled device (CCD) as the photo detector was developed. The OLED was fabricated on a glass plate by photolithography and a vacuum deposition technique. The OLED produced a green luminescence with a peak emission at 512 nm and a half bandwidth of 55 nm. The maximum external quantum efficiency of the OLED was 7.2%. The emission intensity of the OLED at 10 mA/cm(2) was 13 μW (1.7 mW/cm(2)). The fluorescence detection system consisted of the OLED device, two band-pass filters, a five microchannel poly(dimethylsiloxane) (PDMS) microfluidic device and a linear CCD. The fluorescence detection system was successfully used in a flow-based enzyme-linked immunosorbent assay on a PDMS microfluidic device for the rapid determination of immunoglobulin A (IgA), a marker for human stress. The detection limit (S/N=3) for IgA was 16.5 ng/mL, and the sensitivity was sufficient for evaluating stress. Compared with the conventional 96-well microtiter plate assay, the analysis time and the amounts of reagent and sample solutions could all be reduced.