Effect of Intrapulmonary Shunt on the Accuracy of Partial CO_2 Rebreathing Technique for Cardiac Output Monitoring

Ling LIU,Haibo Qiu,Yi YANG,Fengmei GUO
DOI: https://doi.org/10.3760/j.issn:1671-0282.2005.09.006
2005-01-01
Abstract:Objective To evaluate the accuracy of partial CO_2 rebreathing technique,and the effect of intrapulmonary shunt. Methods In 26 mechanically-ventilated patients, cardiac index was measured by noninvasive cardiac output monitor (NICO monitor)based on the partial CO_2 rebreathing method (CI_NICO),and by pulmonary artery catheter based on thermodilution method (CI_TD). Intrapulmonary shunt (Qs/Qt) was calculated by the shunt formula and NICO monitor. Accroding to the Qs/Qt calculated by the shunt formula, patients were divided into low Qs/Qt group (Qs/Q15%) and high Qs/Qt group (Qs/Qt15%). Results In low Qs/Qt group (n=16), no significant difference was found between the Qs/Qt calculated by NICO monitor and the shunt formula. In this group, a good correlation was found between CI_NICO and CI_TD (r=0.91,P0.01), and the difference between CI_NICO and CI_TD was not significant [(2.24±0.42) L/(min·m2) vs. (2.25±0.40) L/(min·m2),P=0.84]. In high Qs/Qt group (n=10), the Qs/Qt calculated by NICO monitor was markedly lower than the shunt formula (P0.01). A significant correlation was also found between CI_NICO and CI_TD (r=0.81,P0.01) in this group, with CI_NICO significantly lower than CI_TD [(2.56±0.64) L/(min·m2) vs. (2.87±0.64) L/(min·m~2), P0.01].Conclusions Correlation between CI_NICO and CI_TD is significant. Qs/Qt is an important factor affecting the accuracy of CI_NICO.