Influence of continous veno-venous hemofiltration therapy on PiCCO monitering data

Qing HE,Zhe FENG,Jing-hua WANG,Zhi-gang CHANG,Pu-xian TANG,Tao-tao LIU,Ya-lin LIU
DOI: https://doi.org/10.3760/cma.j.issn.0254-9026.2010.07.015
2010-01-01
Abstract:Objective To explore the influence of continuous veno-venous hemofiltration (CVVH) therapy on cardiac index (CD, global end-diastolic volume index (GEDI) and extravascular lung water index (ELWI) monitored by Pulse-indicated continuous cardiac output (PiCCO). Methods The 12 critically ill patients with acute renal failure were selected consecutively from department of intensive care unit (ICU) in Beijing Hospital. The patients who received CVVH therapy were monitored by PiCCO plus device. The hemodynamic measurements were performed during the process and interruption of CVVH. Pared t-test was used to analyze the values of CI, GEDI and ELWI. Results A total of 48 groups of data from 12 patients were analyzed. There were no significant differences in CI [(4.75±0.93) L ? min(-1) · m(-2)vs. (4.69±0.89) L · min(-1) · m(-2)], GEDI [(780.60±109.30) ml/m2 vs. (784.75± 106.20) ml/m2] and ELWI CC11.61±3.45) ml/kg vs. (11.54±3.56) ml/kg] between the interruption and process of CVVH, as obtained by the pared t-test (all P>0.05). Conclusions The CVVH therapy has no effect on the accuracy and reliability of PiCCO measurement of CI, GEDI and ELWI.
What problem does this paper attempt to address?