Accuracy of Intraoral Scanned Digital Model with Metal Brackets

Wang Shuo,Liu Yan,Zhou Yanheng,Li Jing,Zhang Da,Wang Linchuan,Liu Xiaomo
DOI: https://doi.org/10.3760/cma.j.issn.1674-5760.2016.02.003
2016-01-01
Abstract:Objective To evaluate the difference of linear measurements between intraoral scanned digital models before and after metal bracket bonding.Methods Intraoral scans (Trios,3Shape,Denmark) on the upper arches were applied to 15 orthodontic patients before and after metal brackets bonding in Department of Orthodontics,Peking University School and Hospital of Stomatology.Three dimensional images were obtained from this scanner.Distance between incisors and the first molar(L1_6/R1_6);width of the first molar(6_6);width of the canine(3_3);clinical crown height of the upper incisors(h1 1/h21)were measured twice.The interclass correlation(ICC) was applied to evaluate the reproducibility.Paired T tests were used to evaluate measurement differences between the two sets of models.Results The values of ICC were above 0.90,indicating the measurement methods were reliable.No statistically significant differences were found in distance between the incisors and the first molar(L1_6;R1_6);width of the first molar(6_6);width of canine (3_3);and clinical crown height of upper right incisors(h11)(P> 0.05).The clinical crown height of left upper incisors(h21)was higher in the models with metal brackets(9.65±0.62) mm than the models without metal brackets(9.51±0.68) mm (P< 0.05).Conclusions Accuracy of intraoral TRIOS scanned digital models with metal brackets can meet the clinical needs on hard tissue's scanning.