Comparison of Central Anterior Chamber Depth Measurements by Pentacam, IOL-Master and A-Scan

LIU Min,ZHAO Jun,ZHAO Hua,QIANG Jun,LIU Li
DOI: https://doi.org/10.3760/cma.j.issn.1006-4443.2011.04.016
2011-01-01
Abstract:Objective To explore the differences of central anterior chamber depth (ACD) determined by Pentacam, IOL-Master and A-Scan and to evaluate the agreement of the three instruments. Methods A total of 126 eyes of 63 subjects aged (25.09 ± 1.29) with spherical Equivalent-(- 5.19 ± 2.04)D were determined ACD on both eyes with Pentacam, IOLMaster and A-Scan. Data were used to analyze the difference of ACD between the three methods. Results The mean value of ACD was (3.718± 0.282) mm for Pentacam and (3.723± 0.295) mm for IOL-Master and (3.584± 0.261) mm for A-Scan. Pentacam and IOL-Master showed a high corre lation and there was no significantly difference. Compared the A-Scan with the Pentacam and IOL-Master measurements significantly misestimate ACD by a mean of (0.133± 0.025) mm (P <0.05), (0.139±0.020) mm (P <0.05),respectively and there was significantly significant difference. Conclusions Pentacam and IOL-Master can be interchanged for ACD, but A-Scan and Pentacam or IOL-Master cannot be interchanged fully.
What problem does this paper attempt to address?