A Method of Measuring Anterior Chamber Volume Using the Anterior Segment Optical Coherence Tomographer and Specialized Software

Ningli Wang,Bingsong Wang,Gaoshou Zhai,Kun Lei,Lan Wang,Nathan Congdon
DOI: https://doi.org/10.1016/j.ajo.2006.11.051
IF: 5.488
2006-01-01
American Journal of Ophthalmology
Abstract:PURPOSE: To describe and evaluate a new method for measuring anterior chamber volume (ACV).DESIGN: Observational case series.METHODS: The authors measured ACV using the anterior chamber (AC) optical coherence tomographer (OCT) and applied image-processing software developed by them. Repeatability was evaluated. The ACV was measured in patient groups with normal ACs, shallow ACS, and deep ACs. The volume difference before and after laser peripheral iridotomy (LPI) was analyzed for the shallow and deep groups.RESULTS: Coefficients of repeatability for intraoperator, interoperator, and interimage measurements were 0.406%, 0.958%, and 0.851%, respectively. The limits of agreement for intraoperator and interoperator measurement were -0.911 mu l to 1.343 mu l and -7.875 mu l to -2.463 mu l, respectively. There were significant ACV differences in normal, shallow, and deep AC eyes (P < .001) and before and after LPI in shallow AC (P < .001) and deep AC (P = .008) eyes.CONCLUSIONS: The ACV values obtained by this method were repeatable and in accord with clinical observation.
What problem does this paper attempt to address?