Comparison of Corneal and Lens Density Measurements Obtained by Pentacam and CASIA2 in Myopes

Yijia Xu,Yuhao Ye,Yiyong Xian,Lingling Niu,Xingtao Zhou,Jing Zhao
DOI: https://doi.org/10.1186/s12886-023-03199-3
IF: 2.086
2023-01-01
BMC Ophthalmology
Abstract:To investigate the agreement between Pentacam and CASIA2 in the evaluation of corneal densities (CDs) and lens densities (LDs) in myopes. Fifty-three patients (106 eyes) underwent comprehensive ophthalmologic examinations. CDs and LDs were measured using Pentacam and CASIA2, respectively, based on the grayscale percentage of the obtained images. Agreement between Pentacam and CASIA2 was evaluated using the consistency intraclass correlation coefficient (ICC) and represented using Bland-Altman plots. Compared to Pentacam, CASIA2 showed significantly higher CD and LD values in all measured zones. The ICC of the average CD and LD measured by the Pentacam and CASIA2 were 0.726 and 0.757, respectively. The ICC values of all corneal zones and lenses were above 0.7, except for the measurement of the cornea in the 0–2 mm zone (0.455), suggesting good consistency between the two devices, whose results were of different levels of linear correlation. Bland-Altman plots showed mean percentages of 3.93
What problem does this paper attempt to address?