Development of Large Area Silicon Semiconductor Detectors for Use in the Current Mode

XP Ouyang,ZF Li,GG Zhang,YK Huo,QM Zhang,XP Zhang,XC Song,HY Jia,JH Lei,YC Sun
DOI: https://doi.org/10.7498/aps.51.1502
IF: 0.906
2002-01-01
Acta Physica Sinica
Abstract:Large area silicon semiconductor detectors for use in the current mode, with their dimensions of phi40, phi50 and phi60 mm, their depletion thickness of 200-300 mum, have been developed. Their performance measurements have been made, which indicate that the developed detectors can satisfactorily meet the needs in expectation. Compared with the detectors commercially available on the market, our large PIN detectors can serve both as reliable and efficient high-resolution devices for nuclear counting experiments, as well as monitors of high-intensity radiation fields in the current mode under a bias of 100-1000 V, while the detectors commercially available are only for the counting use.
What problem does this paper attempt to address?