A study on verifing a general performance detection system for intrinsic performance of SPECT

姚树林,耿建华,陈英茂,徐白萱,刘亚强,马天予
DOI: https://doi.org/10.3969/J.ISSN.1672-8270.2017.05.007
2017-01-01
Abstract:Objective:To verify the reliability of a general slit phantom and general quality control (QC) detection system for single-photon emission computed tomography (SPECT). Methods: The main intrinsic performances of three set of SPECT scanners from two manufacturers were tested by using the general QC detection system of SPECT and the general slit phantom (self-test). And then, for the purpose of comparison, these tested results were compared with the results that were tested by special slit phantom and software from manufacturers (manufacturer-test), respectively.Results: For intrinsic spatial uniformity, a good consistence was found between the self-test and one of the manufacturer-tests, while for another result of manufacturer-test, the consistency was not good. Further analysis showed that the self-test was more reliable than manufacturer-test. For intrinsic spatial resolution, all of the self-test results were consistent with the results of manufacturer-test. However, for the intrinsic spatial linearity, a poor consistence was found between the self-test results and the results of manufacturer-test, but the self-test results existed in between the two manufacturer-tests.Conclusion: The general slit phantom and general QC detection system for SPECT is reliable and the results are more accurate.