Influence of Phantom on the Verification Results of High Contrast Resolution of Medical CT Images

Pu ZHANG,Jia-yu SUN,Jiao LI,Jie SUN,Guang-zhi WANG
DOI: https://doi.org/10.3969/j.issn.1674-1633.2016.11.007
2016-01-01
Abstract:In this paper, the influence of phantom on verification results of high contrast resolution of medical CT images in daily Quality Assessment (QA) tests was analyzed and discussed. And the CTP528 module’s design principle was analyzed under the premise of dissecting its detailed structure. Meanwhile, an Modulation Transfer Function (MTF) estimation method based on SD was used according to CTP528 to calculate the high contrast resolution verification results. The experimental data demonstrated that, under the invariable conditions of scanning mode, slice thickness, reconstruction algorithm and FOV (Field of View), the variation of tube current or tube voltage would not bring apparent change on the verification results of high contrast resolution. Moreover, in addition to the performance of CT equipment itself, the design structure, the place means as well as the machining error of the phantom would also bring a certain degree of influence on the verification results of high contrast resolution. Therefore, as the CT verification phantoms have always been considered as the symbol of “standardization”, however, they still have many shortcomings. Phantoms could play their roles in QA testing better if they were fully understood.