CT image quality evaluation in the age of deep learning: trade-off between functionality and fidelity

Kai Yang,Jinjin Cao,Nisanard Pisuchpen,Avinash Kambadakone,Rajiv Gupta,Theodore Marschall,Xinhua Li,Bob Liu
DOI: https://doi.org/10.1007/s00330-022-09233-0
IF: 7.034
2022-11-10
European Radiology
Abstract:To quantitatively compare DLIR and ASiR-V with realistic anatomical images. CT scans of an anthropomorphic phantom were acquired using three routine protocols (brain, chest, and abdomen) at four dose levels, with images reconstructed at five levels of ASiR-V and three levels of DLIR. Noise power spectrum (NPS) was estimated using a difference image by subtracting two matching images from repeated scans. Using the max-dose FBP reconstruction as the ground truth, the structure similarity index (SSIM) and gradient magnitude (GM) of difference images were evaluated. Image noise magnitude (σ), frequency location of the NPS peak (f peak ), mean SSIM (MSSIM), and mean GM (MGM) were used as quantitative metrics to compare image quality, for each anatomical region, protocol, algorithm, dose level, and slice thickness.
radiology, nuclear medicine & medical imaging