Study of Temperature Influence on CsI(Tl) Detector in High Temperature Gas-cooled Reactor Pebble-flow Detection System

Jianchao LU,Ximing LIU,Xin WAN,Zhifang WU
DOI: https://doi.org/10.7538/yzk.2017.youxian.0345
2018-01-01
Abstract:In order to ensure the output signal stability of pebble-flow detection system for pebble-bed high temperature gas-cooled reactor (PB-HTGR) during the projection acquisition process,the analysis for dark current stability of detector module is required.In this paper,two sets of CsI(Tl) detector modules were randomly selected and a corresponding controlling and measuring device for temperature and dark current was designed.In addition,the dark current characteristic of the detector module and the controlling requirement were analyzed.Experimental results show that the change of dark current with temperature by each detector unit is approximately exponential.Besides,the dark current fluctuate of the 32nd unit reaches 8 pA when the temperature fluctuate is ±2 ℃ at the room temperature.Therefore,a specialized temperature controlling system for the detection system is to be adopted inevitably.Combining the experimental result with dark current controlling requirement,an optimal temperature controlling range is (18-22) ℃,and the corresponding temperature controlling precision is (±0.92-±0.21) ℃.