Repeatability of a primary thermometer and calibration of resistance thermometers between 5 K and 25 K
Haiyang Zhang,Yaonan Song,Xiangjie Kong,Bo Gao,Mark Plimmer,Laurent Pitre
DOI: https://doi.org/10.1088/1681-7575/ad87f5
2024-10-19
Metrologia
Abstract:This work describes improvements to our Single Pressure Refractive-Index Gas Thermometry (SPRIGT) system [Metrologia 57 (2020) 065006], where a new experimental procedure for Constant Refractive-Index Gas Thermometry (CRIGT) was proposed and implemented, and after 2.5 years new values of T–T 90 obtained in the temperature range from 5 K to 25 K. Comparison results indicate that our SPRIGT system is robust with excellent repeatability, reliability, and stability. Not only is the uncertainty in thermodynamic temperature T still below 0.17 mK, but also good reproducibility of T 90 (temperature of the 1990 International Temperature Scale), T and T—T 90 is observed with values of 21 μK, 28 μK and 35 μK, respectively. Furthermore, we have implemented direct dissemination of thermodynamic temperature T by establishing our own thermodynamic temperature wire scale under the new International System of Units (SI) from 5 K to 25 K. The remarkable agreement further enhances our confidence on SPRIGT for implementing the redefined kelvin below 25 K, disseminating thermodynamic temperature T, and ultimately partaking in an international comparison of thermodynamic temperature via highly stable resistance thermometers.
instruments & instrumentation,physics, applied