Evaluation on Methods for the Morphological Observation of Polysulfone Hollow Fiber Membrane in Hemofilter

董兴刚,陈江华,陈万涛,何强,寿张飞,周晓健,万澎波,杨毅,李春霞
DOI: https://doi.org/10.3969/j.issn.1001-7119.2008.05.009
2008-01-01
Abstract:The polysulfone hollow fiber membrane (PHFM)is the basic material to build the renal tubule assist device (RAD). In this study, we compared different methods to observe the structure of PHFM, including general microscope, immunofluorescence staining(IF), laser scanning confocal microscope(LSCM), scanning electronic microscope(SEM)and atomic force microscopy (AFM). Our results showed that the interior and outer walls of PHFM could be observed using IF, LSCM, SEM and AFM, but not the general microscope. Besides SEM, which is more complicated to use, IF and AFMare also reliable tools. The IF is the simplest way to get clear pictures of PHFM. Three-dimentional high-resolution pictures can be obtained with AFM. In general, the immunofluorescence staining and the atomic force microscopy are the most useful methods to observe the polysulfone hollow fiber membranes in hemofilter.
What problem does this paper attempt to address?