Evaluation of matrix effects for serum ALT and AST measurements

ZHENG Song-bai,WANG Jian-bing,HUANG Xian-zhang,MA Yan,ZHUANG Jun-hua,XU Ning,ZHOU Hua-you,CHEN Cha
DOI: https://doi.org/10.13602/j.cnki.jcls.2011.05.024
2011-01-01
Abstract:Objective To evaluate the matrix effects of measurement for alanine transaminase(ALT) and aspartate transaminase(AST) in serum by using 7 materials of external quality assessment(EQA) and 4 commercial reference materials in 9 testing systems.Methods According to the Clinical and Laboratory Standards Institute(CLSI) EP14-A2 guideline and the reference methods of ALT and AST measurement,the comparative methods without pyridoxal 5-phosphate were developed.The routine measurement methods of 9 testing systems were the evaluated.The levels of ALT and AST in all fresh frozen serum samples,EQA materials and commercial materials were measured with the routine methods and the comparative methods,and the matrix effects of non-serum samples were evaluated simultaneously.Results For ALT measurements,one EQA material showed the matrix effects in all the routine analytical systems,and for AST measurements two EQA materials showed no matrix effect in all the routine analytical systems.Different matrix effects of other materials were observed in different routine analytical systems.Conclusion The accuracy and the comparability of ALT and AST measurement are affected by matrix effect.The influence of matrix effect on the traceability of measuring results of clinical samples should be considered.