Clinical Experience of Transvenous Implantable Cardioverter Defibrillator

王景峰,伍卫,张燕,聂如琼,周淑娴,孔敏仪
DOI: https://doi.org/10.3969/j.issn.1007-2659.2004.05.005
2004-01-01
Abstract:To summarize the technical experience of the implantation of implantable cardioverter defibrillator(ICD) by a subclavicular puncture approach ,clinical data of the implantation of ICD, the induction of ventricular fibrillat ion, the detection of defibrillation threshold(DFT), and the programming of the parameters of ICD were analyzed. Results: ICD was successfully implanted by the subclavicular puncture approach i n ten patients,6 cases had a history of frequent syncope.Clinical documented ven tricular fibrillation was noted by ECG during syncope occurred in 2 patients and sustained ventricular tachycardia or fibrillation was induced in the other 8 ca ses. 8 of them suffered from coronary artery disease(one patient combining of Brugada s yndrome); One patient suffered from dilated cardiomyopathy; The other one was primary ventricu lar fibrillation. 5 of 10 patients took aminodarone.All patients were successful induced ventricular fibrillation and were successful defibrillated. The average time for the operat ion was 92±27 minutes. DFT≤20 J, shock path impedance 41.2±15.3 ohms, the amplitude of R wave was 16.3±6.6mV. T here were no complication in all patients. Conclusion:Implantation of ICD by the subclavicula r puncture approach is simple and reliable. Oral aminodarone before ICD implanted might not effect inducing ventricular fibrillation during procedure.
What problem does this paper attempt to address?