Clinical observation of cardiac resynchronization therapy defibrillator without routine defibrillation testing

Yan ji,Xu jian,Fan xizhen,Sun xianlin,An chunsheng
DOI: https://doi.org/10.3969/j.issn.1005-0272.2010.02.002
2010-01-01
Abstract:Objective To explore the necessity of routine defibrillation testing in the implantation of cardiac resynchronization therapy defibrillator (CRT-D). Methods From July 2008 to October 2009, all patients with approved indications of CRT-D were selected. The lead of left ventricular was implanted according to standard methods, right ventricular lead was in apical, and right atria lead was in atrial appendage by subclavian vein puncture or phlebotomy of cephalic vein. Defibrillation testing was not performed in patients with satisfactory parameters, 20 J was thought to be the defibrillation threshold (DFT). Pacemaker was optimized by echocardiography at follow-up, ventricular tachyarrhythmia was recorded and analysed at the same time. Patients received optimized anti-heart failure drugs during follow-up. Results 22 patients were enrolled (19 with sinus rhythm, mean age 63.1±10.5 years), 20 of which were male. Aetiology of heart failure was dilated cardiomyopathy in 77% of the patients. The mean left ventricular end-diastolic diameter was 74.5±13.8mm, left ventricular ejection fraction was 28.6±3.8%, QRS duration was 143.8±18.5ms preoperative. Ventricular tachyarrhythmia were recorded in 4 patients, one of them experienced 28 attacks because of hypokalemia, all could be identified and converted to sinus by CRT-D effectively. Two of four patients could be defibrillated by shock after unsuccessfully antitachycardia pacing (ATP). The remaining one could convert to sinus by ATP. Conclusions Patients with satisfactory parameters could terminate ventricular tachyarrhythmia effectively without routine defibrillation testing by CRT-D.
What problem does this paper attempt to address?