A Double-Sided Radiating Flat-Panel X-ray Source Using ZnO Nanowire Field Emitters

Daokun Chen,Yuan Xu,Guofu Zhang,Zhipeng Zhang,Juncong She,Shaozhi Deng,Ningsheng Xu,Jun Chen
DOI: https://doi.org/10.1016/j.vacuum.2017.08.015
IF: 4
2017-01-01
Vacuum
Abstract:A double-sided radiating flat-panel X-ray source device has been developed. The device has a diode configuration composed of an anode target of a tungsten thin film and field-emitter arrays of ZnO nanowires prepared on glass substrates. The X-ray radiation comes from both transmission and reflection of X-rays induced by electron bombardment. When operated at 24 kV with an anode current of 3.16 mA, radiation dose rates of similar to 200 mGy/min and similar to 210 mGy/min were recorded positioning at 4 cm away from the front and back sides of the vacuum-sealed flat-panel X-ray source, respectively. Clear X-ray images with high contrast were taken by the radiation from both sides of the device. The reported device has great potential in making compact X-ray radiography and irradiation systems. (C) 2017 Elsevier Ltd. All rights reserved.
What problem does this paper attempt to address?