The Effectiveness of Automatic Cleaning and Disinfection Appliance for Gastroscope in Removing of Hepatitis B Virus

曹梅娟,邵爱仙,顾青,方英,范骏
DOI: https://doi.org/10.3969/j.issn.1006-2785.2004.06.004
2004-01-01
Abstract:Objective To investigate the effectiveness of automatic cleaning and disinfection appliance for gastroscope contaminated with hepatitis B virus. Methods After used for patients with positive serum HbsAg, gastroscopes were cleaned and disinfected by automatic disinfection machine. Samples were collected before and after disinfection. Immunoassay was used to detect HBsAg, PCR(polymerase chain reaction) techniques were used to detect HBV-DNA. Results Fifty samples were collected before and after disinfection, among which 15 samples with positive HBsAg and 12 samples with positive HBV DNA before disinfection, all turned to negative after treatment. Conclusion Automatic mechanical disinfection can effectively remove HBsAg and HBV-DNA within 3 min, and PCR technique for HBV DNA is more reliable and sensitive than immunoassay for HBsAg in determining the HBV contamination.