Factors Affecting Sterilization of HBV on Dental Hand Piece

SHI Lei,NIU Guang-liang,TIAN Jing-hua,ZHAO Hui,LIU Ya,XIE Yao
DOI: https://doi.org/10.3321/j.issn:1005-4529.2006.06.021
2006-01-01
Abstract:OBJECTIVE To investigate the factors influencing the effects of disinfection of HBV on dental hand piece. METHODS After polluted with HBV positive serum,the dental hand piece was disinfected with sterilizer at the condition of 134℃ for 5 minutes according different treatment procedures.The samples were then taken and detected HBsAg by the automatic enzyme immune analyzer using the third generation EIA agent.The efficacy of disinfection of HBV was presented by the reaction of HBsAg detection. RESULTS If the dental hand piece was sterilized directly not cleansed using clean water after pollution and dried in room temperature,there were 96.87% of the samples being HBsAg positive,and the polluted but dried hand piece,even cleansed there were 87.5% of positive result, even if prolonged the time of sterilization and the frequencies of vacuumization,the HBsAg positive rate of the sample was still up to 56.25%.But if the hand piece was irrigated and cleansed immediately after pollution and sterilized,all samples were HBsAg negative. CONCLUSIONS Whether or not cleansed immediately is the most important factor influencing the efficacy of hand piece sterilization after pollution.