Stress Distribution under Compression Loading on Butter-fly Fractured Fragment in Middled Part of Fresh Human Tibia Bonded by Medical Graded Hard Tissue Adhensive

吕波,屠重棋,裴福兴,陈孟诗,刘雷
DOI: https://doi.org/10.3969/j.issn.1002-0179.2004.01.018
2004-01-01
Abstract:Objective:To study the machanical compression property of the medical grade hard tissue adhensive bonding butter-fly fractured fragment.Method:A new adhensive,octyl-a-cyanoacrylate,was tested in 5 fresh human tibias. A 90° butter-fly fractured fragments in the middle part of the tibia was made by bandsaw. The biomachnical property under compression loading was tested before and after osteoectomy as the contral group. After bonded by octyl-a-cyanoacrylate adhensive,the machnical property of the tibia under compression loading was tested again as experiment group. Result:Under compressive loading, the slope of the compression curve of the tibia before osteoectomy was higher than that of the bonded tibia; and the slope of the compression curve of the bonded tibia was higher than the unbonded tibia. The compression loading distributed on the bonded fractured fragment was 47.2% of the total compression load 3.75KN.Conclusion:Medical grade hard tissure adhensive bonded butter-fly fractured fragment of the middle part of tibia could improve the compression stress of the tibia bone. Before the failure of the adhensive,the fragment could tolerate 47% of compression load.The result indicated that using medical grade hard tissue adhensive to bond butter-fly fracture fragment of the middle part of tibia could fulfill the biomachanical requiremento of human body.