Determination of Nickel in Hydrogenated Oil

张智宏,李凤生,朱毅青
DOI: https://doi.org/10.3969/j.issn.1001-1803.2004.05.016
2004-01-01
Abstract:The coloring reaction of dimethyl glyoxime with nickel in aqueous solution was studied.The conditions for coloring reaction were optimized.Different sequences of adding reagents were compared,and the optimal sequeuce was identified as:standard solution,dimethyl glyoxime,ammonium persulfate and sodium hydroxide.Following such sequence,the coloring reaction was quick and stable with wide linear range(0.1 mg·L~(-1)~10 mg·L~(-1)) in the standard curve and correlation factor of (0.999 74.)Three kinds of pretreatment methods were compared for the determination of nickel in hydrogenated oil.The appropriate pretreatment was identified as to dissolve the oil sample in chloroform and to extract with acid.
What problem does this paper attempt to address?