Preparation and Performance Characterization of Novel Solid Ph Sensing Electrodes

叶瑛,邬黛黛,黄霞,陈鹰,陈慧明,杨灿军,陈宁华
DOI: https://doi.org/10.3969/j.issn.1004-1699.2003.04.026
2003-01-01
Abstract:Ir/IrO 2 pH electrode is prepared with oxidation method in LiCO 3 melt and paired with a novel Ag/AgCl reference. Characterization results show that responded potential difference between the work and reference electrodes is in linear relation with pH value in the full range of pH, with a slope of-62.429 mV/pH, intercept of 607.97 mV, correlation coefficient R 2=0.993. Impact of temperature on the response signal of the pH sensor is in accord with Nernst equation, easy to be corrected. Salinity's relation with the signal is also linear one, so that pH value can be worked out from response potential in the medium with known salinity, or in the condition that the salinity is measured simultaneously.