Resolution for in Situ Ph Sensing on Seafloor: Ir/IrO2 Electrode with Nafion Shielding Membrane

Ying Ye,Xia Huang,Canjun Yang,Ying Chen,Wei Zhao,Daidai Wu
DOI: https://doi.org/10.3969/j.issn.1004-1699.2006.06.037
2006-01-01
Abstract:Ir/IrO2 electrode is vulnerable to dissolved sulphur anion, which is commonly found in hydrothermal fluid at middle ocean ridge (MOR) and cold seep in gas hydrate area on continental slope. Naked Ir/IrO2 electrode with Ag/AgCl as reference responds linearly with pH standard series, almost the. same as predicted by Nernst Equation. The same pairs in a buffered solution (pH=12.6) with 2x10(-4) mol/L Na2S failed to yield stable signal. Instead there was strong false signal at the beginning, followed by a bumping curve in the rest time of scanning. To overcome the disturbance from sulphur anion, the Ir/IrO2 electrode was coated with Nafion, a kind of proton selective substance produced by DuPont. Nafion coated Ir/IrO2 electrode performed well in pH standard series, comparable with the uncoated counterpart. In the buffered solution with 2x10(-4) mol/L Na2S, the responding signal is stable. This confirms that the interference from sulphur anion is shielded out. Solid state sensing pairs of Ir/IrO2 and Ag/AgCl are suitable for pH measurement and long term monitoring on seafloor, for both hot and cold venting systems.