The influence of different diluents on the human chorionic gonadotrophin detected by electrochemilumines-cence immunoassay

Xiaomei MO,Huahui LI,Pingping WEI,Yantuan LI
DOI: https://doi.org/10.3969/J.issn.1672-6790.2013.06.014
2013-01-01
Abstract:Objective To explore the impact of the different diluents on the serum β-hCG by Electrochem-iluminescence immunoassay.Methods A total of 27 serum samples from 3 ~8 weeks of pregnancy were collected and divided them into 3 groups by hCG concentration:low concentration group(hCG < 1 000mIU /ml),medium concentration group (hCG 1001 ~10 000 mIU /ml),and high concentration group(hCG >10 000mIU /ml).Di-luted with Elecsys universal dilution,0.9% sodium chloride injection solution,distilled water in 1:10 dilution,respec-tively.The hCG level was detected by Roche Cobas E601 chemiluminescence analyzer.Results In the low concentra-tion group,the results diluted by saline solution were different from those by Elecsys universal dilution (P =0.042), while there were no differences in the other two groups (P =0.201,P =0.076);There were significant difference be-tween Elecsys universal dilution and distilled water among three groups (P =0.000).Conclusion The influence of diluents on the β-hCG concentration detected by electrochemiluminescence immunoassay is obvious.Saline solution and distilled water are not available for low concentration group,while saline solution could be substitute for diluents in medium and high concentration group.