[Methodological Study on Magnetic Enzymic Immunoassay for Detecting Free Hcgbeta Subunit].

Zhen-shi Liu,Hao Zhou,Hai-sheng Chen,Jing Xie,Xin Lu,Xiang-liang Yang
DOI: https://doi.org/10.3321/j.issn:1007-8738.2006.03.038
2006-01-01
Abstract:AIM:To establish a novel magnetic enzyme immunoassay (MEIA) for detecting human choriogonadotropin free beta subunit (hCGbeta).METHODS:Two monoclonal antibodies (mAb) were used for conjugating with FITC and with alkaline phosphatase(AP) respectively, which incorporated magnetic solid phase separation. Magnetic beads were coupled with sheep anti-FITC antibody as solid phase, and phenolphthalein monophosphate was used as substrate to set up MEIA for detecting hCGbeta.RESULTS:The sensitivity of hCGbeta MEIA kit reached 0.1 IU/L. The intraassay variation and inter-assay variation was 8.5% and 14% respectively, with the average recovery rate of dilution of 92.5%. The kit showed no cross reactivity to LH, TSH and FSH, while a cross reactivity of 1.2% to intact hCG at concentration of 1000 IU/L. The time of efficacy of hCGbeta MEIA kit was longer than 14 months.CONCLUSION:hCGbeta MEIA kit is better than hCGbeta radio immunoassay and ELISA kit, which can provide a high-qualitative and cheap hCGbeta kit for market.