Magnetic Microparticle Chemiluminescence Immunoassay Method for Determination of Estriol in Human Urine

Zheng Guo-Jin,Fang Lu-Qiu,Chen Hui,Li Zhen-Jia,Ying Xi-Tang,Lin Jin-Ming
DOI: https://doi.org/10.3724/sp.j.1096.2011.00062
2012-01-01
Abstract:A magnetic microparticle chemiluminescence immunoassay method was developed for the determination of estriol in human urine. The sample of estriol and horseradish peroxidase (HRP) labeled estriol with fluorescein isothiocyanate (FITC) labeled rabbit anti-estriol antibodies were mixed in homogeneous system. The reaction product in homogeneous system can be combined onto the goat anti-FITC antibody-coated magnetic particles. After the magnetic field separation of bond and free, the light-emitting substrate was added to magnetic particles. Under the optimization conditions, this methods can be used to detect estriol in a linear range of 1-100 mu g/L, detection limit of 0.25 mu g/L, relative standard deviation less than 14% and between-relative standard deviation less than 7%. The method has been applied for the determination of estriol in real samples, and the results are satisfactory.
What problem does this paper attempt to address?