Reagentless amperometric immunosensor for human chorionic gonadotrophin based on direct electrochemistry of horseradish peroxidase.

Jin Chen,Feng Yan,Zong Dai,Huangxian Ju
DOI: https://doi.org/10.1016/j.bios.2004.10.023
IF: 12.545
2005-01-01
Biosensors and Bioelectronics
Abstract:A novel amperometric immunosensor for determination of human serum chorionic gonadotrophin (HCG) was constructed by immobilization of HCG with titania sol–gel on a glassy carbon electrode and the direct electrochemistry of horseradish peroxidase (HRP) labeled to HCG antibody (HRP-anti-HCG). The morphologies of the HCG membrane were characterized to be chemically clean, porous and homogeneous. HRP-anti-HCG was functionally conjugated with the immobilized HCG after incubation in phosphate buffer (PBS) containing HRP-anti-HCG. A direct electron transfer of HRP with a rate constant of 1.35±0.40s−1 was observed at the HRP-anti-HCG-HCG/titania sol–gel membrane modified electrode in 0.1M PBS pH 7.0. With a competitive mechanism the differential pulse voltammetric peak current of the immobilized HRP decreased linearly with an increasing HCG concentration from 2.5 to 12.5mIU/ml in the incubation solution. The HCG immunosensor showed a detection limit of 1.4mIU/ml, a good accuracy and acceptable precision and reproducibility with an intra-assay CV of 4.7% at 5.0mIU/ml and an inter-assay precision of 8.1% obtained at 10mIU/ml. The biosensor displayed a good stability in a storage period of 30 days.