Influence of Dilution Medium and Method on the Determination of Serum Estradiol Level with Chemiluminescence Immunoassay

BAO An-yu,LI Yan,SU Han-wen
DOI: https://doi.org/10.3969/j.issn.1006-1703.2011.04.019
2011-01-01
Abstract:To explore the influence of dilution medium and method on detection of serum estradiol concentration using chemiluminescence immunoassay,the precision and linearity of the analysis system was evaluated and verified before the study.Twenty-seven serum samples which the estradiol concentration was within the linear range were selected for the dilution experiment.Each sample was divided into four parts,one was measured for estradiol with automatic dilution using SIEMENS Centaur with assorted dilution medium D9,while the other three parts were diluted manually by assorted dilution medium D9,mixed low-level estradiol serum and normal saline respectively,and then measured by SIEMENS Centaur.The diluent fold was five for all parts.The results of diluted estradiol were collected and analyzed,and they were compared with the original values.Paired t-test was applied to observe the difference of dilution group and original group.The relative bias,correlation coefficient and linear regression were analyzed as well.The results showed that the diluted estradiol level was correlated well with the original value in all of the four groups.No significant difference was found in each dilution group and original group(P>0.05).However,the lowest bias was obtained in the group of automatic dilution and assorted dilution medium.The best and most appropriate measurement of high level estradiol was using instrument automatic dilution mode and with the assorted dilution medium.