Performance of paramagnetic chemiluminescence microparticle immunoassay for determination of serum folic acid

Chong-wen AN,Hai-xia LI,Li-li JIAO,Hong-yun YANG,Ya-jun WANG
DOI: https://doi.org/10.16155/j.0254-1793.2014.12.021
2014-01-01
Abstract:Objective:To evaluate the performance of paramagnetic chemiluminescence microparticle immunoassay (CMIA) for serum folic acid.Methods:Performance verification tests for precision,accuracy,anti-interference,analytical measuring range (AMR),clinical reportable range (CRR),biological reference interval and stability at different temperatures and illumination conditions were conducted according to the clinical and laboratory standards institute (CLSI) EP5-A2,EP15-A2,EP7-A2,EP6-A,C28-A3 c guidelines.Results:The precisions of within-run and between-run were lower than the stated specification by manufacturers when the concentration of folic acid was 2.68-28.17 nmol · L-1.The accuracy test results of SRM 1955 met the allowable range of the target value.The results of EQA samples (K-C and LN5-B) also met the CAP calibration and validation/linear evaluation error limits standards,and met linear verification when folic acid concentration was from 1.26 to 30.83 nmol · L-1 The 95% verification interval also contained the assigned value.The relative deviation was less than the standard of external quality assessment by national center for clinical laboratory (TEa:target value ± 30%).However.The overall results were relatively low,especially for the low level and the maximum relative deviation was up to-28.44%.Anti-interference showed that the triglyceride (TG) concentration ≤ 20 mmol · L-1,bilirubin (Bil) concentration ≤400 pmol · L-1,and vitamin C mass concentration ≤2 g · L-1 had no significant interference on the folic acid detection system (CMIA).AMR validation showed that the best fit equation was linear equation polynomial.There was linear relationship when the concentration of folic acid was 0-45.31 nmol · L-1.The upper limit of CRR was 6796.5 nmol · L-1,and the maximum dilution was 150 times.Reference interval validation showed that the overall level of folic acid in reference individuals were lower than that in Americans.The levels of folic acid in males were lower than those in females.Serum folic acid should not be stored more than 2 days at normal temperatures.Conclusion:Performance of CMIA for detection of serum folic acid met laboratory requirements.Stability was better when samples were stored sealed,away from light and at low temperatures.The laboratory should establish or verify the reference interval according to regional dietary differences.
What problem does this paper attempt to address?