Methodological evaluation of microbiological assay in detection of plasma folate

Liu Jian-meng
2011-01-01
Abstract:Objective:In order to obtain accurate and reliable test results,microbiological assay on 96-microtitre plates of folate in plasma is evaluated.Methods: According to the guidelines provided by the AOAC,plasma folate concentration of SRM and pooled plasma by microbiological assay were measured and got evaluation indicators of this assay including accuracy,precision and parallelism test results etc.Results: The measuring range of the standard curve is from 0.025 μg/L to 0.05 μg/L;The relative error of testing SRM with 2 replicates each is 2.07%~5.36%;Recovery is 107%~120%;Intraassay and interassay precision are 4.32%~7.86% and 3.05%~8.08%,respectively;The method showed acceptable reproducibility,as evidenced by HORRAT values of 0.25~0.32;Parallelism Test results are considered within acceptable limits;The data from the chemiluminesence immunoassay and Microbiological assay were highly correlated(r=0.96).Conclusion: Microbiological assay on 96-well microtitre plates of Plasma folate showed acceptable accuracy and precision.It is a reliable and pratical method for large-scale epidemiological studies,considering its low cost,simple preparation for samples and simple equipment requirements.