Quantitative analysis of HBsAg:comparison of two assay diluents

XING Yan-qing,JI Fen-zhi,WANG Lei,YANG Fang
DOI: https://doi.org/10.16505/j.2095-0136.2012.04.010
2012-01-01
Abstract:Objective To compare two diluents in the HBsAg quantitative detection.Methods A total of 96 specimens with HBsAg over 250 IU/ml were included in the study.ARCHITECT HBsAg detection kit was used for further dilution analysis to compare two diluents:solution dilution in the kit(A method) and saline dilution(B method).Results Data from 91 cases were included in the statistical analysis.The detection range of methods A and B were(2.72-4.92)and(2.52-4.97)lg IU/ml,respectively;the medians were 3.83 and 3.81 lg IU/ml,respectively.There was no significant difference between diluents A and B(P=0.712).The correlation coefficient of the two methods was 0.969 and with high consistency by Bland-Altman analysis;the mean of difference between method A and B was close to 0.Methods A and B also showed very high correlation at different HBsAg levels.Conclusions Normal saline can be used as the diluents in the detection of HBsAg when using the ARCHITECT kit.