Dosimetry of a China-made 192ir Wire Source

何昆仑,王所亭,高焱章,周凯欣,冯宁远,邱学军
DOI: https://doi.org/10.3760/cma.j.issn.2095-2848.2001.02.024
2001-01-01
Abstract:Objective To evaluate the dose releasing rate and the homogeneity of dose distribution of the Chine-made intravascular 192Ir wire source and to provide experimental data of animal study and clinical use. Methods Dosimetry of the 192Ir wire source was measured by Kodak X-omat V film in differents directions and in different exposure lengths.Dose releasing rate of our 192Ir wire source was calculated according to AAPM TG No.60 and Monte Carlo model,and compared with the results by traditional brachytherapy planning system(AAPM TG No.43) Results 192Ir wire source provided with a well-distributed dose around the source,no striking asymmetry was observed.Thirty two percent of dose rate was overestimated by traditional brachytherapy planning system(AAPM TG No.43)comparing with Monte Carlo method. Conclusions The China-made 192Ir wire source is reliable for intravascular radiation.It is an alternative method to measure the dose distribution with Kodak X-omat V film.The dose rate of 192Ir wire source can be estimated by traditional brachytherapy planning system (AAPM TG No.43) and Monte Carlo method.This study may provide a fundation for practicing intravascular radiation in China.