Dosimetry characterization of $^{32}$P intravascular brachytherapy source wires using Monte Carlo codes PENELOPE and GEANT4

Javier Torres,Manuel J. Buades,Julio F. Almansa,Rafael Guerrero,Antonio M. Lallena
DOI: https://doi.org/10.1118/1.1637970
2003-11-22
Abstract:Monte Carlo calculations using the codes PENELOPE and GEANT4 have been performed to characterize the dosimetric parameters of the new 20 mm long catheter based $^{32}$P beta source manufactured by Guidant Corporation. The dose distribution along the transverse axis and the two dimensional dose rate table have been calculated. Also, the dose rate at the reference point, the radial dose function and the anisotropy function were evaluated according to the adapted TG-60 formalism for cylindrical sources. PENELOPE and GEANT4 codes were first verified against previous results corresponding to the old 27 mm Guidant $^{32}$P beta source. The dose rate at the reference point for the unsheathed 27 mm source in water was calculated to be $0.215 \pm 0.001$ cGy s$^{-1}$ mCi$^{-1}$, for PENELOPE, and $0.2312 \pm 0.0008$ cGy s$^{-1}$ mCi$^{-1}$, for GEANT4. For the unsheathed 20 mm source these values were $0.2908 \pm 0.0009$ cGy s$^{-1}$ mCi$^{-1}$ and $0.311 \pm 0.001$ cGy s$^{-1}$ mCi$^{-1}$, respectively. Also, a comparison with the limited data available on this new source is shown. We found non negligible differences between the results obtained with PENELOPE and GEANT4.