Primary application investigation of dose verification using measured bremsstrahlung X-ray spectra of 6 MV medical linear accelerator

张松柏,赵洪斌,黄斐增,张怀苓,胡家臣,韩树奎,包尚联
DOI: https://doi.org/10.3969/j.issn.1005-202X.2003.04.003
2003-01-01
Abstract:With the measured X-ray energy spectra of a 6 MV medical linear accelerator produced by Beijing Medical Equipment Institute by the attenuation method and analysis decomposition algorism- the PDD- OAR curves of the accelerator were calculated- which were compared with directly measured data of PDD- OAR curves for same accelerator. The results showed that the attenuation method and our decomposition algorism is reliable-which can be used for verification for radiotherapeutical treatment plan.