Comparison between Monte Carlo simulation and measurement of beam quality and output factor with flattened and flattening filter-free of linear accelerator

Hai-zhen YUE,Yi-bao ZHANG,Zhuo-lun LIU,Jian GONG,Fan JIANG,Hui ZHANG,Hao WU
DOI: https://doi.org/10.3969/j.issn.1005-202X.2016.06.002
2016-01-01
Abstract:Objective To simulate the flattened and flattening filter-free (FFF) modes of 6 MV photon beams emitted from TrueBeam accelerator by using Monte Carlo method,and to compare the calculated beam quality and output factor with the measured results.Methods Beamnrc and Dosxyznrc programs were used to model the linear accelerator head.And the dosimetric data of two modes under reference conditions were calculated and output for various field sizes.The relative deviations between the calculated beam quality of various field sizes and the measurement were calculated,and the absolute values of relative deviations were statistically analyzed.On the beam central axis of various field sizes,the output factors were calculated at 10 cm depth under the water surface.The absolute differences between the calculated output factors and the measurement were statistically analyzed.Results The absolute values of the relative deviation of 6 MV and 6FFF beam quality were respectively (0.459 ± 0.462)% and (0.486 ± 0.300)%,and those of 6 MV and 6FFF output factor were respectively (1.315± 1.868)% and (0.904±1.214)%.Conclusion For the proposed model,the relative deviations between the calculated beam quality and output factor and the measurement were small,so the proposed model can be applied for clinical dosimetric investigations.
What problem does this paper attempt to address?