Comparison analysis of energy spectra between flattened and flattening-filter-free Monte Carlo models for photon beam

Haizhen YUE,Yibao ZHANG,Jian ZHANG,Zhuolun LIU,Qiaoqiao HU,Fan JIANG,Hao WU
DOI: https://doi.org/10.3969/j.issn.1005-202X.2018.06.003
2018-01-01
Abstract:Objective To compare and analyze the energy spectra of 6 MV flattened and flattening-filter-free photon beam in the air medium and standard water phantom at specific depth. Methods BEAMnrc program was used to model Varian TrueBeam accelerator gantry head, including flattened model and flattening-filter-free model. The phase space files of (40×40) cm2field at SSD of 110 cm in air medium and standard water phantom were calculated. The energy spectrum distribution in different regions of the irradiation field were compared and analyzed with the use of BEAMDP program. Results For the irradiation field of (40×40) cm2at SSD of 110 cm in air medium, with the enlargement of statistical regions, the low energy part in the energy spectrum of flattened model increased, while that of flattening filter-free model decreased. For the standard water phantom at depth of 10 cm, large differences were found in the energy spectrum of two models with or without scatter, mainly in the area less than 0.511 MeV. The low energy part in the energy spectrum of flattened model in different locations decreased in crossplane direction, while the flattening filter-free model showed opposite trend. Compared with electron and position electron, the photon had great effects on the energy spectrum in the same irradiation field. Conclusion This study provides basis for further clinical dosimetric data measurement and calibration of medical linear accelerator.