Abstract:A CMOS-compatible, aluminum nitride (AlN) based, dual-electrode bimorph piezoelectric micromachined ultrasonic transducer (pMUT) array has been developed. Under an input 5V ac driving voltage, a high acoustic intensity of 30-70mW/cm 2 up to 2.5mm deep into tissues can be generated for therapeutic medical applications. The low driving voltage is readily applicable for battery-powered low intensity pulsed ultrasound (LIPUS) medical devices for fracture and tissue healing applications. The fabrication also introduces a novel method using PECVD SiO 2 as the barrier layer between AlN films to increase the breakdown voltage and manufacturing yield. Results on a prototype array have shown the highest intensity per voltage squared, per number of pMUTs squared, per piezoelectric constant squared (i.e. I n =I/(VNd 31 ) 2 ) among all reported pMUT arrays.