The Conceptual Design of Interferometer/polarimeter System on HL-2M.

Y. Zhou,H. X. Wang,Y. G. Li,Y. Li,W. J. Chen,J. Yi,Z. C. Deng
DOI: https://doi.org/10.1063/1.4959911
IF: 1.6
2016-01-01
Review of Scientific Instruments
Abstract:HL-2M is a new medium-sized tokamak with major radius of 1.78 m, minor radius of 0.65 m, and aspect ratio of 2.8 and will be finished soon. In the conceptual design, a double-pass, horizontal view, multi-channel far-infrared (FIR) laser interferometer/polarimeter system (at 432 μm) is proposed to simultaneously measure the phase change and the Faraday rotation for density and current profile reconstruction. A vertical CO2 dispersion interferometer system (with wavelength of 10.6 μm and 5.3 μm) is designed to measure the line density, which will mainly be used for density feedback. The performance of the system is expected to meet the requirement of the measurements on HL-2M, with phase accuracy ∼0.1°, time resolution ∼1 μs, spatial resolution ∼7.2 cm, and density measurement range of 1 × 10 18-1 ×1021 m-3.
What problem does this paper attempt to address?