Optical Path Design of Phase Contrast Imaging on HL-2A Tokamak

Qiyun Cheng,Yi Yu,Shaobo Gong,Min Xu,Tao Lan,Wei Jiang,Boda Yuan,Yifan Wu,Lin Nie,Rui Ke,Ting Long,Dong Guo,Minyou Ye,Xuru Duan
DOI: https://doi.org/10.1088/2058-6272/aa8d64
2017-01-01
Abstract:A phase contrast imaging (PCI) diagnostic has recently been developed on HL-2A tokamak. It can diagnose plasma density fluctuations with maximum wave number of 15 cm(-1) and wave number resolution of 2 cm(-1). The time resolution reaches 2 mu s. A 10.6 mu m CO2 laser is expanded to a beam with a diameter of 30 mm and injected into the plasma as an incident beam, injecting into plasma. The emerging scattered and unscattered beams are contrasted by a phase plate. The ideas of optical path design are presented in this paper, together with the parameters of the main optical components. The whole optical path of PCI is not only carefully designed, but also constructed on HL-2A. First calibration results show the ability of this system to catch plasma turbulence in a wide frequency domain.
What problem does this paper attempt to address?