Calibration of Phase Contrast Imaging on HL-2A Tokamak

Y. Yu,S. B. Gong,M. Xu,C. J. Xiao,W. Jiang,W. L. Zhong,Z. B. Shi,H. J. Wang,Y. F. Wu,B. D. Yuan,T. Lan,M. Y. Ye,X. R. Duan
DOI: https://doi.org/10.1088/1748-0221/12/10/c10005
2017-01-01
Abstract:Phase contrast imaging (PCI) has recently been developed on HL-2A tokamak. In this article we present the calibration of this diagnostic. This system is to diagnose chord integral density fluctuations by measuring the phase shift of a CO2 laser beam with a wavelength of 10.6 mu m when the laser beam passes through plasma. Sound waves are used to calibrate PCI diagnostic. The signal series in different PCI channels show a pronounced modulation of incident laser beam by the sound wave. Frequency-wavenumber spectrum is achieved. Calibrations by sound waves with different frequencies exhibit a maximal wavenumber response of 12 cm(-1). The conversion relationship between the chord integral plasma density fluctuation and the signal intensity is 2.3 x 10(13) m(-2)/mV, indicating a high sensitivity.
What problem does this paper attempt to address?