432-Μm Laser’s Beam-Waist Measurement for the Polarimeter/interferometer on the EAST Tokamak

Z. X. Wang,H. Q. Liu,Y. X. Jie,M. Q. Wu,T. Lan,X. Zhu,Z. Y. Zou,Y. Yang,X. C. Wei,L. Zeng,G. S. Li,X. Gao
DOI: https://doi.org/10.3938/jkps.65.1215
2014-01-01
Journal of the Korean Physical Society
Abstract:A far-infrared (FIR) polarimeter/interferometer (PI) system is under development for measurements of the current-density and the electron-density profiles in the EAST tokamak. The system will utilize three identical 432-μm CHCOOH lasers pumped by a CO2 laser. Measurements of the laser beam’s waist size and position are basic works. This paper will introduce three methods with a beam profiler and several focusing optical elements. The beam profiler can be used to show the spatial energy distribution of the laser beam. The active area of the profiler is 12.4 × 12.4 mm2. Some focusing optical elements are needed to focus the beam in order for the beam profiler to receive the entire laser beam. Two principles and three methods are used in the measurement. The first and the third methods are based on the same principle, and the second method adopts an other principle. Due to the fast and convenient measurement, although the first method is a special form of the third and it can only give the size of beam waist, it is essential to the development of the experiment and it can provide guidance for the choices of the sizes of the optical elements in the next step. A concave mirror, a high-density polyethylene (HDPE) lens and a polymethylpentene (TPX) lens are each used in the measurement process. The results of these methods are close enough for the design of PI system’s optical path.
What problem does this paper attempt to address?