High three‐dimensional accuracy of component placement and lower limb alignment using a robotic arm‐assisted system and gap‐balancing instrument in total knee arthroplasty

Emanuele Diquattro,Robert Prill,Mikhail Salzmann,Francesco Traina,Roland Becker
DOI: https://doi.org/10.1002/ksa.12088
2024-02-29
Knee Surgery Sports Traumatology Arthroscopy
Abstract:Purpose It was hypothesized that robotic arm‐assisted total knee arthroplasty (RA‐TKA) using additionally a gap‐balancing instrumentation will show high accuracy in executing the planning in femoral and tibial component placement throughout the range of knee motion (ROM) during TKA surgery. Methods Prospectively collected data were analysed for patients undergoing RA‐TKA. A cruciate retaining cemented design was implanted using the MAKO® robotic system. Lower limb alignment at 0°, 30°, 45°, 60° and 90° of flexion was recorded at the beginning of surgery and finally after implantation of the components. A ligament tensioner was inserted after tibial precut to measure the extension and flexion gap, and final component placement was planned based on 3D CT images. Femoral and tibial component placement was measured in all three planes. Results A total of 104 patients were included (mean age 69.4 ± 9 years; 44 male, 60 female). The difference in component placement after planning and final implantation showed less valgus of 0.7° ± 1.4° (p
surgery,orthopedics,sport sciences
What problem does this paper attempt to address?